Basic Parameters:
|
Testing Method |
Chromatography/ion-exchange liquid chromatography |
|
Testing Item |
Glycated hemoglobin HbA1c |
|
Testing Scope |
3%-18% |
|
Testing Parameters |
Precision (CV) <3% Accuracy (V) <1.5% |
|
Testing Time |
4 minutes (including reporting) |
|
Sample Type |
Venous blood (EDTA anticoagulation), Peripheral finger blood |
|
Sample Volume |
5 ul whole blood for each sample |
Functional Parameters:
|
Photometer |
415mn LED integral flow colorimeter |
|
Sampling Mode |
Full auto 25 position sample turntable (20 samples) position, 2 quality control position, 1 emergency position, 1 cleaning position, 1 zero position. |
|
Testing Mode |
Auto test for batch, single test for emergency |
|
Calibration |
Optional selection of 2 points or 4 points calibration, optional selection of manual or automation, equipped with 1 set of weight high and low level calibrators |
|
Thermostatic control |
Column and reagent thermostatic control 25 +- 0.5 |
|
Reagent matching |
One piece of column and reagent consumable / 300 pieces of testing sample |
|
Reagent allocation |
Each standard package includes A, B, C, D eluent, calibrator, hemolytic agent, pump tubing |
Output Parameters:
|
Display |
320 x240 LCD Graphic display, displaying real-time testing curve |
|
Printer |
58mm Thermal printer, printing testing curve and report |
|
Report output |
IFCC concentration value, NGSP area percentage, ADAG average clucose |
|
Data storage |
1000 tests report (including testing curve) |
|
Communication interface |
RS232 communication interface, connecting to HIS/LIS system |
Working Parameters:
|
Power supply |
AC220V 50Hz |
|
Size |
408mm x 390mm x 230mm |
|
Weight |
9 Kg |
|
Working environment |
Temperature 12-30 , Relative humidity <85% |