1.Sterilize temperature ≤ 135 °C (Improved type ≤ 155 °C ),if the
|
temperature is over 135 ° C ,the motor will be damaged.
|
2.Normal bone drill can only hold drill bits less than 6mm,acetabulum
|
reaming drill with big torque,rpm is around 300,canulate drill is used
|
with nail or K wire ,rpm is around 600.
|
3.The battery need to be get out of the pwer tool after surgery. Otherwise
|
the small automatic discharge of the electric will make the life of the
|
battery short.
|
4.The battery cannot be sterilized ,it is not autoclavable.
|
5.The cranial bur must be washed right after the surgery by purger in
|
order to aviod the blood stick in the bur,then we should make it dry.
|
6.The battery space in the handpiece of the tool should not have any
|
water,after sterilization.
|
7.The cranial bur blade is made of medical stainless steel,in order to
|
avoid broken,so it is flexible and not very hard,it is easy to be attrite,
|
we suggest each blade only use for 1 surgery.
|